Main INDEX
Monthly INDEX
PREV
NEXT
User name Lingyan
Log entry time 06:41:20 on January12,2001
Entry number 54419
This entry is a followup to: 54358
Followups:
keyword=HV change for S1, gas and aerogel
The small gain of left arm gas cherenkov channel #9 is because the HV too low, only 2200, while the rest is greater than 2500. So we increased it to 2600.
HV changes
===========
between run 1207 and 1209:
left arm s1_R_4 : from 2050 to 2125
left arm gas adc_9 : from 2200 to 2600
right arm AM adc
between run 1208 and run 1210:
left arm A2 adc
after run 1219 before run 1229:
the hv for aerogle detectors were further adjusted to get better gain.
Now the single photo-electron peaks of A1 and AM are around 100, and those of a2 are around 150-200.