Main INDEX
Monthly INDEX
PREV
NEXT
User name kramerk
Log entry time 04:07:02 on January17,2001
Entry number 54876
Followups:
keyword=CO2 gas shed pressure -74psi
The CO2 pressure reading on the gas shed field shows a negative quantity (-72 psi). Joe and Kevin went down to the shed and noticed the regulators on the line to the hall read zero. The regulators on the bottle side of the solenoid valves read 200-300psi. Looking at the data the gas cerenkov detector (presumably the only thing using CO2) seems to be functioning properly (the peak is similar to previous runs where the pressure was higher). Perhaps this is okay. Lingyan printed out a sample cerenkov spectrum sheet for the right arm and we'll compare future runs to this spectra. In the morning, someone should see if this is okay or not.