Log entry time 02:06:54 on January28,2001
Entry number 55911
keyword=Left arm slow controls.
The slow controls crate for the left arm detector stack has been down since about
5 PM on 01/27. When power was restored after the HV crate swap I assume the
ioc tried to reboot. The server for that ioc does not seem to be allowing the ioc
to connect. Thus the ioc will not finish loading. The DAC outputs will be zero
volts. This means the VDCs had no voltage on them and the discriminator
thresholds on the VDCs were also at zero volts. I'm sure the amplifiers on the
VDCs were just oscillating merrily away.