User name jpchen
Log entry time 11:06:52 on February23,2001
Entry number 56837
keyword=He4 re-cool-down completed
He4 target re-cooldown (with cold-return) started by Joe, was completed
at 10:45 am. The condition now is at 5.8K and 212 psia (read by the tranducer,
the gauge at the gas panel reads 222 psia). Fan is at 25% and HP heater is in PID
with about 50 watts and 4K flow is 34 g/s. Coolant temperatures: ESR supply
is 4.25K, T after JT is 4.6, after HX is 6.4K, and on the target return line is
7.85K and after mixed with spectromter return, it is 5.25K (this one normally should
be below 5K). So it is still a little high. Will take some time to settle.