User name jpchen
Log entry time 11:15:11 on February23,2001
Entry number 56839
keyword=TO needs watch pressure/temperature and heat load carefully
To target operators:
Please watch the target temperature and pressure carefully. Since the operating
pressure is around 220-230 psia an dthe vent pressure is 260 psia, there is not
much room for fluctation. Make sure you always have enough HP heater power
to replace the maximum beam heating power and the fan heating power (adjust
J-T valve to increase coolant to increase HP heater power (since it is in PID).
Also always make sure it is in PID control.