User name Reitz
Log entry time 23:53:45 on February26,2001
Entry number 57043
This entry is a followup to: 56726
keyword=bpm/harp calibration pedestals
analysing the bpm-pedestal Run 2211 from Feb26-2001 leads to
the following values for the Struck7510 adc
2086 2084 2084 2084 2079 2074 2072 2071
a gauss was used to fit the peak, the sigma was always <= 3.2
for comparison from Oct02-2000
2084 2082 2082 2082 2077 2072 2070 2069
from May07-2000
2081 2080 2078 2078 2076 2071 2068 2067
But in various databases the following values are usually given:
2105 2104 2098 2098 2090 2085 2085 2086
analysing the harp-scans with these different sets
of pedestals does not lead to significant changes in the bpm-offsets
and/or bpm-slopes (the offsets are changing by less than
15 um, the slopes by less than 3 %)