Main INDEX
Monthly INDEX
PREV
NEXT
User name jpchen
Log entry time 16:31:01 on February28,2001
Entry number 57180
Followups:
keyword=Target operators, please adjust coolant flow (J-T valve) slowly
Notice to cryotarget operators:
Again, please adjust J-T valve SLOWLY. Even decreasing the heat load, do it slowly. Watching the ESR inlet temperature and the coolant flow (A_4K is senstive to
ESR changes) when you make adjustment. Cryo-group send me this plot, showing
the Hall A 15K flow increased from 9g/s to over 16g/s at around 4 am and then decreased from over 15.5 g/s to 5.5g/s in short time. This could be the cause for
the ESR fluctuation sometime later.