    User name R. Michaels

    Log entry time 10:36:37 on March19,2001

    Entry number 59027

    keyword=scaler overflows (negative charge)

    The long runs at high currents leads to scaler overflows.
    E.g. run 2993 which lasted 137 minutes reports average
    current -52 uA and T3 rate -174 kHz. I didn't check, but
    I'm pretty sure the negative number is from overflow and
    the recovery is to look at the 4-second updates of scalers
    in datafile. You will see them ramp up, then jump to negative
    and then ramp up from there again.