Main INDEX
Monthly INDEX
PREV
NEXT
User name R. Michaels
Log entry time 10:36:37 on March19,2001
Entry number 59027
keyword=scaler overflows (negative charge)
The long runs at high currents leads to scaler overflows.
E.g. run 2993 which lasted 137 minutes reports average
current -52 uA and T3 rate -174 kHz. I didn't check, but
I'm pretty sure the negative number is from overflow and
the recovery is to look at the 4-second updates of scalers
in datafile. You will see them ramp up, then jump to negative
and then ramp up from there again.