User name Jack
Log entry time 15:54:36 on April16,2001
Entry number 60652
keyword=S2 and RICH
Pete and I had a little confusion about which PMT signal was bad.
I could not find a problem with the analog signal from S2 R4. I do not know what
the problem was. The connector on S2 L4 needs to be replaced sometime. It is
showing signs of wear and tear, but was still "good".
The PMT bases have incorrect labels on some of them. Be careful when checking
cabling and reporting problems.
The "leak tight" freon system for the RICH is apparently not. Water is condensing
on the top of the chiller and running off. Freon is leaking out the sides of the chiller
unit. This all is running down over the cables and stuff below the freon system.
The cable supports are corroding, the supports bolted to the I-beam with beam
clamps.