There was problem with Harp channles 4, 5 and 6.
We called in Rick Ganzales. First we found that Ch 4 encode
was switched off and ch's 5,6 preamps were switched off.
After turned them on, ch 4 and 6 worked fine but ch 5 still had
a problem: in up scan, you got two lines. We tried with a spare
card, then ch 5 ok, but ch's 1 and 4 had problem. After tried swap
cards and push in connectors, all channles working now.