User name averett
Log entry time 13:18:33 on June 5,2001
Entry number 61965
keyword=beam position summary
After moving the beam spot around on the BPM's by:
(x,y)=0,0 2,0 -2,0 0,2 0,-2mm
we see the harp scans agree with this movement. The SPOT software is reporting a movement of only 1mm in each case. OTR is showing about 1.3 mm of movement.
Conclude that we can trust Harp and BPM, but not SPOT at this time. Will ask R. Michaels to recalibrate SPOT.