Main INDEX
Monthly INDEX
PREV
NEXT
User name K. McCormick
Log entry time 19:23:09 on June 7,2001
Entry number 62383
keyword=Access to check target lifter
Since the accelerator is down, I went in with the Radcon guy to check the vertical target lifter. I used the local control box and ran to the upper and lower limit. The positions for the limits agreed with the positions I noted a few weeks ago to .78 mm for the upper limit and less than .32 mm for the lower limit. This number is only
reproducible to ~.5 mm because of the actuation of the mechanical limit switch.
The point of this was to see if the target frame has slipped. From these numbers,
I don't think the outer frame has. Maybe the inner one with the target ladder
has. That will be checked Monday.