Log entry time 19:09:34 on June 8,2001
Entry number 62554
keyword=Target - Holding Field changed in the 180 degree Config.
I made two adjustments to the field direction in the 180 direction. One the field direction was incorrect. There was a wrong sign in the initialization code. The old transverse field current was : 0.245 A and the new is -0.0795 A. Also the field magnitude was too high. This has to do with an adjustment I make to do the NMR in the right field range. This adjustment is different in the 180 degree direction than in the 0 degree direction. It might take one more small adjustment to get the field correct.