Main INDEX
Monthly INDEX
PREV
NEXT
User name Armstrong
Log entry time 14:59:20 on June 9,2001
Entry number 62666
keyword=charge asymmetry confusion
We had some beam (for Compton tuning) with BeO target.
We took data at 5muA, and then ran beam charge asymmetry analysis
on the run; qasy gave -410 ppm (deltaQ=80) but the parity asymmetry
gave -638 ppm (BCM1) and -706 ppm (BCM2). We do a timestamp
of the asymmetry with -410 ppm - but now we wonder if the feedback is
on at all, and is it working? We note that the beam 1/2 wave plate is
IN, so that the slope is opposite to that in log entry 62548 - so did MCC
update the slope? Is the feedback program running at MCC?
Also, the timestamp is off by 2 hours on that GUI... Why?
Separate aside - we plugged back in the DAQ sync cable before these
runs, so we are back to the Parity DAQ controlling the other 2 DAQs...