    User name xfzhu

    Log entry time 01:14:30 on June11,2001

    Entry number 62865

    keyword=NMR polarization

    12:00
    beam is off
    turn off 3 transverse lasers
    move to pickup coil position
    rotate holding field back to 0 degree
    perform NMR imediately after lasers were off: r=7.85 P=41.15%
    ( the calibration constant used above is 5.24%/mV, which should be used when 3 lasers are on , with Oven temperature at 170 degree. )
    Another calibration constant is 5.79%/mV without lasers on and with the oven temperature at 165 degree, probably 20 minutes after lasers are off .
    The actual calibation constant for the above NMR measurement should be between the above 2 constants.