User name R. Michaels
Log entry time 11:43:21 on June11,2001
Entry number 62871
This entry is a followup to: 62444
keyword=L-arm T1 scaling fixed
It was noticed for a few days (halog 62444) that the L-arm copy
of the trigger 1 (T1) signal was over-counting. In today's access
it was found that a Phillips 757 logic fanout on R-arm produced an
after-pulse of 135 mV 20 nsec after trailing edge; this was sent to L-arm
and was apparently enough to fire logic modules there (with some
probability). Solution was to avoid the 757 and instead use a LeCroy
623B discriminator with 210 mV threshold as a logic fanout.
I don't think there are any serious implications for experiment, as both
the R-arm scaling and the R-arm trigger supervisor T1 signals looked ok.
All other scaling and TS signals were checked on scope today. Ringing
is less than 30 mV.