Main INDEX
Monthly INDEX
PREV
NEXT
User name Zolnierczuk
Log entry time 10:59:37 on June12,2001
Entry number 62891
Followups:
keyword=TARGET NMR/EPR
9:30 Transverse Laser #1,#2 and #3 off
Target moved to Pickup Coil position
9:37 Field rotation 270 -> 0
9:38 Longitudinal Laser #6 on
9:55 EPR DONE delta_nu = 50 kHz (POL.~40%)
10:00 NMR DONE 7.52 mV -> POL. 42.2 %
NOTE 1: x-fit is a little to high
NOTE 2: I used cal.const for 1 laser that I calculated as follows
(5.79 - cc. for 3 lasers, 5.24 cc for one laser)
5.60%/mV = 5.79 - 1/3*(5.79-5.24)
Laser #6 off
10:05 Field rotated from 0 -> 270 deg
Target moved to pol 3He position
10:15 Transverse Lasers (#1, #2 and #3) ON