During Moeller measurements we had a charge asymmetry and
used that to cross calibrate the Moeller apparatus helicity sign versus
the spectrometers. Gene saw +3 +/- 1e-3 asymmetry (he can give more
accuracy later), while I get from scalers:
R-arm +2950 +/- 120 ppm (agrees w/Moeller, also in sign)
L-arm +3000 +/- 134 ppm (" " ")
Strangely, Parity DAQ gives no result because the signal is indistinguishable
from the previously recorded pedestal. For this current 0.6 uA I expected a
signal 2% higher than pedestal, but maybe something is nonlinear. Should
check later at higher current.
Also would be interesting to see what is the sign of Trigger asymmetry
in event DAQ of spectrometer, but there is no target now, must check later.