User name E.Chudakov
Log entry time 07:12:41 on June18,2001
Entry number 63650
This entry is a followup to: 63637
keyword=HV tripped along with the alarm IOC
IOCs for HV alarm were rebooted. It occured that HV for the shower detector (Right Arm)
was OFF. Is was also indicated by the data analysis for the run 1636.
The crate was off at least since 5:00. The run 1635 (finished at 4:55) showed
normal spectra for the shower detector, though one should check the last
period of this run, since the HV crate might have tripped then. The runs
1636 and 1637 (a short one) have no shower detector in the right arm.
The problem was fixed before the run 1638 started.