User name X. Zheng, W. Korsch
Log entry time 17:31:53 on June21,2001
Entry number 64044
keyword=EPR - 38.96%
Performed EPR at 15:00 hrs
Conditions: Cell position: pol 3He, laser #6 on, beam: off, trans. grad coil: 6.00 A,
long. grad. coil: -3.00 A (reverse)
EPR Sweep: failed
Called Xiaochao.
15:50 hrs: Next sweep after adjustment of electronics. Conditions as above.
EPR Sweep: worked
P = 38.96%
frequency shift down: -46.91 +/- 1.36 KHz
frequency shift up: 46.68 +/- 1.64 KHz
polarization down: -38.96% +/-1.13%
polarization up: 38.77% +/- 1.36%
It appears that we are running in pretty stable conditions and the effect of
the larger trans. grad coil current is small (if it had an effect at all?).
If there is no change in about four hours (after the next EPR), we might
want to start lowering the long. grad. coil current (from -3.00 A to -2.00 A).
Given the present conditions we should be able to switch to the 180 deg
configuration tomorrow morning.