• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name xfzhu , Karl S., Korsch W.

    Log entry time 11:15:52 on June23,2001

    Entry number 64256

    keyword=EPR and NMR measurement

    9:45 am
    During data taking , with beam on, laser #6 on,
    gradient coil transverse -6.0A (reversed), longitudinal -2.0A (reversed).

    0th order fit:
    -42.4 +/- 2.9 KHz, P = 35.2% +/- 2.4%
    42.6 +/- 2.2 KHz, P = 35.4% +/- 1.8%


    10:10 am
    turned off all the lasers NMR measurement
    x = -3.71 mV
    y = 3.80 mV
    v = 5.3 mV

    10:23 am
    turned on laser #6, EPR measurement with beam on, during data taking.

    0th order fit:
    -41.3 +/- 2.9 KHz, P = 34.3% +/- 2.3%
    40.9 +/- 2.2 KHz, P = 33.9% +/- 1.9%

    the EPR measurements show consistantly depolarization due to AFP. No EPR and NMR till 18:00 pm.