• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name Wolfgang

    Log entry time 20:15:57 on June24,2001

    Entry number 64446

    keyword=Target - EPR - 34.9 %

    19:30 EPR measurements

    Failed twice to lock signal during flip!!!

    Third time it worked, changed Wavetek frequency from 11.39 MHz to 11.41 MHz
    before third sweep.

    Conditions: Laser #6 on, target in Pol 3He Cell position, long. grad. coil: -2.00 A,
    trans. grad. coil: +6.00 A, beam on.

    P = 34.9 %

    0th order fit:

    freq. shifts:
    down: -42.0 kHz +/- 0.8 kHz
    up: 41.8 kHz +/- 0.9 kHz

    pol's:
    down: 34.9 % +/- 0.7 %
    up: 34.8 % +/- 0.8 %

    1st order fit

    freq. shifts:
    down: -41.9 kHz +/- 0.8 kHz
    up: 41.6 kHz +/- 0.9 kHz

    pol's
    down: 34.8 % +/- 0.7 %
    up: 34.5 % +/- 0.8 %

    If we lost ~1% per AFP, we probably had 36.5 % - 37 % before first sweep.

    Maybe we should even wait longer, before we do next EPR????