Log entry time 00:38:00 on Juni27,2001
Entry number 64718
keyword=EPR/NMR , holding field changed to 270 degree
23:50 EPR measurment
Conditions: target in pick up coil position, laser #6 on, beam off ,
trans. grad coil: +5.00 A, long. grad coil: -2.00 A
P = 34 %
0th order fit:
freq. shifts:
down: -41.6 kHz +/- 2.4 kHz
up: 41.6 kHz +/- 2.1 kHz
pol's:
down: 34.6 % +/- 2.0 %
up: 34.6 % +/- 1.8 %
1th order fit:
freq. shifts:
down: -40.6 kHz +/- 2.0 kHz
up: 41.9 kHz +/- 2.1 kHz
pol's:
down: 33.8 % +/- 1.7 %
up: 34.8 % +/- 1.8 %
When EPR measurement finished, I tried to close the EPR window,pressed the "exit" from the pull out window of "File" option, which exit the whole TO screen.vi
24:03 turn off laser #6, NMR measurement
24:05 rotate holding field from 0 degree to 270 degree.
24:12 turn on 3 transverse lasers
24:13 target cell back in beam line position