Main INDEX
Monthly INDEX
PREV
NEXT
User name W. Korsch, J.P. Chen
Log entry time 10:06:28 on June27,2001
Entry number 64769
keyword=Target - EPR - 40 %
9:20 am EPR measurement
Conditions: pol. 3He Cell position, field at 0 deg, laser #6 on, long. grad coil: -2.00 A, trans. grad coil: +5.00 A, beam off (tuning)
P = 40 %
0th order fit:
freq. shifts:
down: -48.2 kHz +/- 3.2 kHz
up: 45.9 kHz +/- 3.9 kHz
pol's:
down: 40.0 % +/- 2.7 %
up: 38.1 % +/- 3.2 %
1st order fit:
freq. shifts:
down: -48.5 kHz +/- 3.2 kHz
up: 47.1 kHz +/- 3.9 kHz
pol's:
down: -40.3 % +/- 2.7 %
up: 39.1 % +/- 3.2 %
Comment: The whole spectrum has a slight slope (drift of B-field -> we moved
from transverse to longitudinal direction just before measurement),
so 1st order fit is more realistic.
Lock-in signal quite noisy.