Main INDEX
Monthly INDEX
PREV
NEXT
User name Wolfgang
Log entry time 18:03:00 on June28,2001
Entry number 64902
keyword=Target - EPR
17:20 EPR measurement
Conditions: position: pick-up, laser #6 on, trans. grad. coil: +5.00 A,
long. grad. coil: -2.00 A, beam: off
P > 42 %
Comment: First spin flip happened very late (effect of integration time?) and
frequency shift did NOT reach minimum.
I still analysed the run, but DON'T trust these numbers:
0th order fit:
freq shift:
down: -51.0 kHz +/- 2.8 kHz
up: 45.2 kHz +/- 2.6 kHz
pol's
down: 42.4 % +/- 3.2 %
up: 40.8 % +/- 3.2 %
Decided not to try another EPR-AFP sweep.
Just to make sure that this is not real, performed a 2nd NMR measurment
and the polarisation is still high (see separate entry).