    User name X. Zheng

    Log entry time 00:29:03 on July 1,2001

    Entry number 65209

    keyword=Target - EPR - 37%

    12:15am, EPR, laser #6 on, during data taking, beam on
    target postion: polarized 3he cell.

    0th order fit:
    down: freq. shift = -44.19 KHz +/- 4.96 KHz, P = -36.70%+/- 4.12%
    up: freq. shift = 45.34 KHz +/- 4.78 KHz, P = 37.66%+/- 3.97%

    1st order fit:
    down: freq. shift = -44.57 KHz +/- 5.19 KHz, P = -37.02%+/- 4.31%
    up: freq. shift = 44.14 KHz +/- 4.81 KHz, P = 36.66%+/- 4.00%

    Note: D2 signal is okay but lock-in has a high fluctuation. Not sure if it is because
    beam on/ diode radiation damage/ electronics problem. They all could cause
    high noise but we don't have enough data to prove any one of them.

    I put the average of 4 (2 fits x 2 flips) numbers in the title.