User name X. Zheng
Log entry time 00:29:03 on July 1,2001
Entry number 65209
keyword=Target - EPR - 37%
12:15am, EPR, laser #6 on, during data taking, beam on
target postion: polarized 3he cell.
0th order fit:
down: freq. shift = -44.19 KHz +/- 4.96 KHz, P = -36.70%+/- 4.12%
up: freq. shift = 45.34 KHz +/- 4.78 KHz, P = 37.66%+/- 3.97%
1st order fit:
down: freq. shift = -44.57 KHz +/- 5.19 KHz, P = -37.02%+/- 4.31%
up: freq. shift = 44.14 KHz +/- 4.81 KHz, P = 36.66%+/- 4.00%
Note: D2 signal is okay but lock-in has a high fluctuation. Not sure if it is because
beam on/ diode radiation damage/ electronics problem. They all could cause
high noise but we don't have enough data to prove any one of them.
I put the average of 4 (2 fits x 2 flips) numbers in the title.