User name Wolfgang, Xf. Zhu
Log entry time 21:08:16 on July 1,2001
Entry number 65334
keyword=Target - EPR - 39 % (large error)
20:20 EPR measurement
Conditions: cell position: pol. 3He, laser #6 on, B field: 0 deg., trans. grad. coil:
+ 5.00A, long. grad. coil: -2.00 A, beam: on (data taking off)
P ~ 39 %
0th order fit:
freq. shift:
down: -47.36 kHz +/- 8.8 kHz
up: 46.5 kHz +/- 11.0 kHz
pol's:
down: 39.3 % +/- 7.3 %
up: 38.6 % +/- 9.0 %
1st order:
freq. shift:
down: 46.7 kHz +/- 8.6 kHz
up: 46.9 kHz +/- 10.0 kHz
pol's:
down: 38.8 % +/- 7.1 %
up: 39.0 % +/- 8.7 %
Comment: It is clear from the error bars that the signal is extremely noisy.
Nevertheless we believe that the central value is not too bad, since we waited
a little longer (20 s or so) before and after the sweep to have enough points
to fit. I had to turn the integration off during the down flip and turned
it on again before the up flip. The down flip is not good, but there is enough
data to fit the bottom spin state.
My suspision is that the diode dies slowly due to radiation damage.
Note: Although I clicked on "print" in the temperature panel, the window was
either never printed or somebody else picked it up. I did wait long enough to
let the temperatures stabalize (RTD6 ~ 172.4 deg C, RTD7 ~ 166 deg C).