User name X. Zheng
Log entry time 15:59:00 on July11,2001
Entry number 65784
Followups:
keyword=Target - Duke EPR 41.6%
6:51 am, EPR for Duke
3 lasers (#4, #6, #7) were used
photodiode signal DC -46.7mV, D2 signal ~ +/- 40 uV from lock-in
gradient coils off
Calibration constant obtained from :
pumping chamber temperature 222 C (from Kevin's entry 65782),
target chamber temperature 50 C,
room temperature density: 9.18 amg,
pumping chamber volume 121.1 cc,
target chamber volume 75.9 cc, (from UVa cell data sheet).
Result:
1st order fit:
down: freq. shift = -49.10 KHz +/- 2.20 KHz, P = -41.65%+/- 1.87%
up: freq. shift = 48.34 KHz +/- 2.19 KHz, P = 40.99%+/- 1.86%
0th order fit:
down: freq. shift = -49.09 KHz +/- 2.25 KHz, P = -41.63%+/- 1.91%
up: freq. shift = 48.48 KHz +/- 2.25 KHz, P = 41.11%+/- 1.91%
AFP measurement is stable: