User name W. Korsch
Log entry time 13:04:23 on July13,2001
Entry number 65951
keyword=Target - NMR, EPR - 48.8 %
12:00 pm Performed a sequence of NMR and EPR measurements:
Conditions: beam off, trans. and long. grad. coils off, target in pick-up coil
position, lasers #5,6,7 on, B field: 0 deg.
First NMR:
P ~< 50.66 % (fitted peak heights too large, so real polarisation lower!!!)
EPR:
P = 48.8 % (average of all four numbers (two from 0th order, two from 1st order))
0th order fit:
freq. shifts:
down: -53.81 kHz +/- 2.2 kHz
up: 52.65 kHz +/- 2.3 kHz
1st order:
freq. shifts:
down: -54.22 kHz +/- 2.2 kHz
up: 53.22 kHz +/- 2.2 kHz
Since I did not have the correct calibration numbers for Tilghman, I extracted
the factor 0.913 from Xiaochao's previous measurement.
Comment: NMR value is clearly too high and EPR constant might be a little
off, but the EPR value should be closer.