User name W. Korsch
Log entry time 19:36:03 on July19,2001
Entry number 66778
keyword=Target - EPR - 41.7%
18:25 EPR measurements , important: see note below
P(initial): 44.0 %
P(final): 41.7% (average of last 2 flips)
Conditions: beam off, target position: pick-up coil, lasers #5,6,7 on, grad. coils off,
B-field: 0 deg.
0th order fit:
freq1: -48.11 kHz +/- 1.75 kHz | Polarisation: -43.93 % +/- 1.6 %
freq2: 46.43 kHz +/- 1.58 kHz | Polarisation: 42.4 % +/- 1.45 %
freq3: -45.66 kHz +/- 1.72 kHz | Polarisation: -41.7 % +/- 1.57 %
freq4: 45.25 kHz +/- 1.78 kHz | Polarisation: 41.33 % +/- 1.63 %
first order fit:
freq1: -48.48 kHz +/- 1.68 kHz | Polarisation: -44.27 % +/- 1.53 %
freq2: 45.73 kHz +/- 1.49 kHz | Polarisation: 41.76 % +/- 1.36 %
freq3: -46.47 kHz +/- 1.76 kHz | Polarisation: -42.44 % +/- 1.61 %
freq4: 45.12 kHz +/- 1.81 kHz | Polarisation: 41.21 % +/- 1.65 %
Comments: While watching for spin flip to occur on target computer screen, noticed
that no spin flip happened!!! Looked at oscilloscope and saw the RF was not
displayed, looked at HP function generator: amplutide = 2.4 V(rms), green LED
on. So that's ok. What was wrong!!! Toggled NMR/EPR switch on capacitor
box again and then RF appeared!! (Switched from NMR to EPR position
before EPR measurement). As soon as RF was on, spin flip happened and
then RF went off -> ended up in wrong spin state!! Started immediately
new sweep to flip spin back. Then I waited a little too long before I turned off
the RF button, i.e. second spin flip happened. Performed another (1/2)
sweep before finally spins ended up in right spin state again.
To summarize: two complete sweeps (four flips) happened in total.
NOTE: NMR/EPR toggle switch is getting flaky -> always watch oscilloscope
and check whether RF is really on.