keyword=EPR Procedure for Tilghman at Transverse Mode
When target is at transverse (270 degree) mode, should
turn off transverse lasers;
rotate field to 0 degree;
turn on 3 longitudinal lasers (#5, #6 and #7);
wait until temperature is stable (15 minutes);
Perform EPR;
turn off longitudinal lasers;
Close epics control windows for longitudinal lasers and open windows for transverse lasers;
rotate field to 270 degree;
turn on three transverse lasers;
continue taking data.
Calibration constant for Tilghman with 1 laser (#6) is:
Polarization(%) = 0.92690144 * freq shift (KHz)