• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name Reitz

    Log entry time 20:12:26 on July25,2001

    Entry number 67578

    keyword=Cleaner Trigger Copies

    observations and changes during todays access:
    (R.M. , B.R.)

    left arm:
    copy of T1 on left arm scalers :
    looks rather clean, at least much cleaner
    than the original copy on the right arm.
    ( we didnt see any clear double pulses on the scope,
    but it is hard to see them, if they apear only at
    the percent level)

    T3 on the left arm scalers:
    looks also clean, but we found some.
    (we saw 8 double pulses during 100 sec (approx 1000 Events,
    but don't know about efficiency of the scope))

    left arm scalers:
    produced a "clean" copy of T3 (using a discriminator as gate
    generator with approx 150 ns width)
    connected it to channel 30 of the normalization scalers.

    connected a copy of the EDTM pulser to channel 31 and 32


    right arm:

    connected a copy of the EDTM pulser to channel 30 and 31
    of the right arm normalization scalers.
    ( here the "clean" copy of T1 was alread connected to channel 32 )