keyword=Target - Results of temperature study
These are the results of the two measurements I did of the temperature this morning:
Pumping Chamber Temp w/ 3 lasers = 230C(503K)
Target Chamber Temp = 43C(316K)
Here are some useful numbers for doing calibrations:
Room Temp. pressure = 8.45 amagats
Pump. Chamber volume = 119.1 cm3
Target Chamber volume = 79.6 cm3
Total volume = 201.6 cm3