User name W. Korsch
Log entry time 12:53:06 on August 5,2001
Entry number 69421
keyword=Late entry: Network/DAQ problems
After the network problems this morning the software person (CC) on call
(Mike Memory ??) came and checked out the network. According to his data logs the
cause of the network problem was due to several (self) reboots of the network switches hallasc-cat and halla-cat located in the Hall A electronic room (middle
room of counting house). It turned out that both switches were connected to
the same UPS and so the expert suggested to remove the UPS (nothing else
was connected to it) and use the power directly from an outlet (he promised to
bring a new UPS tomorrow - we should make sure that it'll happen).
After the reset of the switches the EPICS readout of the right arm spectrometer
still didn't work. The accelerator software person on call was very difficult
to reach (>1 hour response time), but finally showed up around 8:30 am.
Supposedly he checked everything and came to the conclusion that we
should contact J. Gomez (~9:45am).
We also noticed that the right arm DAQ did not collect any data (rates etc
were showing 0 al the time after a run started, left arm was ok). Contacted
Bob Michaels several times and he had suggested earlier already to reboot
the VME crates in the right arm.
Requested access around 10:00 am and reset the VME crate on the lower level
of the detector hut (162, slow controls). EPCIS readout worked again,
but still problems with right arm DAQ (not rates...). Contacted
Bob Michaels again and he started looking into it from home.
The problem was related to the rcServer process. Bob rebooted adaql2
from home and then we started rcServer and coda again.
Took a short pulser run in the right arm (run# 21609, ps8=10) and
everything seemed ok (see entry 69403 by Bob Michaels).
Resumed data taking aroun 11:40 am.