Main INDEX
Monthly INDEX
PREV
NEXT
User name T. Averett, W. Korsch
Log entry time 11:52:03 on August 7,2001
Entry number 69739
keyword=Target -NMR: 40.6%, EPR: 41.5%
11:10 MCC took off beam for RF recovery. NMR/EPR measurements
Conditions: beam off, turned off trans. lasers #1,2,4; move target to pick-up coil
position, rotate field to 0 degrees, NMR measurment, turn on lasers #5,6,7;
wait for temp. to stabilize (8 min), EPR measurement, turn off long. lasers,
move target to pol. 3He position, rotate field back to 270 deg, lasers # 1,2,4 on
again.
NMR: P = 40.6%
EPR: P = 41.5%
1st order fit:
freq1: -44.38 kHz +/- 1.54 kHz | Polarisation: 41.51 % +/- 1.44 %
freq2: 44.31 kHz +/- 1.67 kHz | Polarisation: 41.44 % +/- 1.56 %
Comments: 1st order fit only since B-field was still drifting, Vmod = 0.85 V,
good sweep (a little noisy).