• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name T. Averett, W. Korsch

    Log entry time 11:52:03 on August 7,2001

    Entry number 69739

    keyword=Target -NMR: 40.6%, EPR: 41.5%

    11:10 MCC took off beam for RF recovery. NMR/EPR measurements

    Conditions: beam off, turned off trans. lasers #1,2,4; move target to pick-up coil
    position, rotate field to 0 degrees, NMR measurment, turn on lasers #5,6,7;
    wait for temp. to stabilize (8 min), EPR measurement, turn off long. lasers,
    move target to pol. 3He position, rotate field back to 270 deg, lasers # 1,2,4 on
    again.

    NMR: P = 40.6%

    EPR: P = 41.5%

    1st order fit:
    freq1: -44.38 kHz +/- 1.54 kHz | Polarisation: 41.51 % +/- 1.44 %
    freq2: 44.31 kHz +/- 1.67 kHz | Polarisation: 41.44 % +/- 1.56 %

    Comments: 1st order fit only since B-field was still drifting, Vmod = 0.85 V,
    good sweep (a little noisy).