• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name W. Korsch

    Log entry time 12:04:11 on August 8,2001

    Entry number 69927

    keyword=Target - NMR: 36.4%, EPR: 37.8 %

    11:10 am NMR/EPR measurements

    Conditions: B-field at 0 deg, beam off, cell position: pick-up coil, all grad. coils
    off, last run numbers: 2692/21692

    NMR: all lasers off

    P = 36.4%

    EPR: lasers #5,6,7 on

    P = 37.8 % (average of two 1st order fits)

    1st order fit:
    freq1: -39.57 kHz +/- 1.89 kHz | Polarisation: 37 % +/- 1.77 %
    freq2: 41.3 kHz +/- 1.73 kHz | Polarisation: 38.62 % +/- 1.62 %

    Comments: Only 1st order fit, since B-field was still drifting. Signal noisier,
    Vmod = 0.8 V, should go back to Vmod = 0.7 V next EPR.

    Note: Rotated B-field to 90 degrees after polarization measurements.