Check and sdjusted laser wavelength: (With help from Dave, Kevin and Mike)
Before starting checking, three longitudinal lasers were on.
All longitudinal lasers were pretty good. Lasers #6, #7, do not need adjustment.
#5 adjusted slightly, temperature was set/read=18.0/17.9 adjusted to 18.1/17.9.
Transverse lasers all needed adjustment: (they were off before checking)
Laser T( Set/Read) After Adjustment
#1 13/13 14/14
#2 20.7/21.0 22.1/22.1
#3 26.2/26.2 28.0/27.2
#4 15.0/14.8 15.4/15.0
Then changed to transverse pumping with lasers #1,2,4 on for about 1.5 hours.
Check/and adjusted again:
#1 and #4 unchanged
#2, from 22.1/22.1 to 22.4/22.3
#3, from 28.0/27.5 to 27.9/27.4
Hope these will also help with the transverse pumping efficiency.