    User name Zolnierczuk

    Log entry time 14:21:18 on August24,2001

    Entry number 72163

    keyword=target NMR=41.5%/EPR=42.9%

    target at 0deg, BeO position, long. lasers 5,6,7 on

    13:19 move the target to pickup coil position
    - note afrer the move the cell readings changed
    see t=-54mins on the plot

    13:22 NMR P=42.1%

    BT=-0.1187 A BL=-7.0675 A

    13:30 EPR P=42.9%

    long. lasers 5,6,7 off

    13:34 back to pol3He position, but after a thought decided to take another NMR
    so move to pickup again

    13:40 NMR P=41.5%

    13:45 rotate from 0 - 90 degree and target moved to pol3He
    NOTE: the target l/2 plate still in

    13:49 transverse lasers 1,2 and 4 on

    13:53 ready for beam



    EPR fit details
    ---------------------
    0th order fit
    req1: -45.79 kHz +/- 3.31 kHz | Polarisation: 42.82 % +/- 3.1 %
    freq2: 45.39 kHz +/- 3.1 kHz | Polarisation: 42.45 % +/- 2.9 %


    1st order fit
    freq1: -45.46 kHz +/- 3.28 kHz | Polarisation: 42.51 % +/- 3.07 %
    freq2: 46.28 kHz +/- 3.08 kHz | Polarisation: 43.28 % +/- 2.88 %




    FIGURE 1