User name X. Zheng
Log entry time 01:30:47 on August26,2001
Entry number 72510
keyword=missing entry for EPR @ Aug-25, 8:24am
Aug-25, 8:24am, EPR performed after NMR,
latest run number 3114/31114
conditions: all grad. coils off, beam off, 3 longitudinal lasers on,
target @ pick-up coil position, Vmod 0.75 V -> 0.7 V, RF 16.0V,
sweep/waiting time = 10/24 sec.
P = 41.15%
1st order fit:
down: freq. shift = -44.26 KHz +/- 2.15 KHz, P = -41.39%+/- 2.01%
up: freq. shift = 43.73 KHz +/- 1.95 KHz, P = 40.90%+/- 1.83%
comments:
use lower Vmod to minimize the error bar, at the same time use longer sweeping
time (slower sweeping) to make sure the system will lock in to the resonance even
with this low Vmod=0.7V. (I haven't done any test on sweeping time vs. Vmod
yet, just by assumption this might help).
Suggest next time take EPR at Vmod 0.75V, sweep/wait time = 8/24 sec
(though this entry is late due to halog problem)