Main INDEX
Monthly INDEX
PREV
NEXT
User name Xiaofeng
Log entry time 17:16:59 on August27,2001
Entry number 72722
keyword=Target NMR: 45.6% EPR: 46.2%
16:27 beam off
16:30 move target to pick up coil position
NMR with 3 longitudinal lasers on, P=45.6%
16:35 EPR P=46.2%
target@pick up coil position, holding field@0 degree. gradient coil off.
EPR 0th order fit:
freq1: -49.74 kHz +/- 1.09 kHz | Polarisation: 46.51 % +/- 1.02 %
freq2: 49.46 kHz +/- 1.23 kHz | Polarisation: 46.25 % +/- 1.15 %
1st order fit:
freq1: -49.44 kHz +/- 1.09 kHz | Polarisation: 46.23 % +/- 1.02 %
freq2: 49.11 kHz +/- 1.27 kHz | Polarisation: 45.92 % +/- 1.19 %
comments: Vmod=0.85Vpp, Time(wait/sweep)=16/6second, shape is not good. Should use 24/8 instead.
16:43 turn off 3 longitudinal lasers ,and rotate holding field from 0 to 270 degree
16:50 target@270 degree, turn on transverse lasers(#1,2,4) and ready for beam