Main INDEX
Monthly INDEX
PREV
NEXT
User name Wolfgang, Xf. Zhu
Log entry time 18:07:41 on September 4,2001
Entry number 73332
keyword=Target - EPR: 42.6%
17:20 EPR P= 42.6%
Target: Virginia-I@ polarized 3He position, beam off ,laser #5,6,7 on.
Gradient coil: 2A longitudinal. 5A transverse
0th order fit:
freq1: -43.82 kHz +/- 2.07 kHz | Polarisation: 42.93 % +/- 2.03 %
freq2: 43.02 kHz +/- 2.22 kHz | Polarisation: 42.14 % +/- 2.18 %
1st order fit:
freq1: -43.8 kHz +/- 2.03 kHz | Polarisation: 42.91 % +/- 1.99 %
freq2: 43.26 kHz +/- 2.21 kHz | Polarisation: 42.38 % +/- 2.17 %
17:31 EPR P= 41.8%
Target: Virginia-I@ polarized 3He position, beam off ,laser #5,6,7 on.
Gradient coil: 1.6A longitidinal. 5A transverse
0th order fit
freq1: -42.84 kHz +/- 2.34 kHz | Polarisation: 41.97 % +/- 2.29 %
freq2: 42.32 kHz +/- 2.56 kHz | Polarisation: 41.45 % +/- 2.51 %
1st order fit:
freq1: -42.65 kHz +/- 2.41 kHz | Polarisation: 41.78 % +/- 2.36 %
freq2: 42.9 kHz +/- 2.4 kHz | Polarisation: 42.03 % +/- 2.35 %
Comments: Cooling jet flow: 13.8 l/min, Vmod = 1.00 V, good flips, a little
noisy.
Reduced long. grad coil currrent to 1.6 A to check the effect of
energized HRS magnets on masing (note: cell position in pol. 3He this time).
Did not see any masing at 1.6A, but polarization was 42% and not 45% as
this morning. Still went back to 2.00 A to be safe.
Why was polarization lower than this morning? Still recovering from
target/radiation survey and pumping in transverse mode for most of the
afternoon.