User name X. Zheng
Log entry time 17:52:28 on September 6,2001
Entry number 73643
keyword=Target - EPR 39.4% with one laser
17:19, turn on laser #6 after NMR,
17:24, EPR performed, (program started at 17:21), P = 39.4%
conditions: cell positon = pickup coil,
photodiode signal DC=-123mV, D2 signal ~ 15 uV,
Vmod = 1.0 V at the beginning, then changed to 1.2Vpp in the middle because the first flip was very slow. Second (up) flip looks better with 1.2 Vpp
sweep/waiting time = 6/30 sec.
Suggested new EPR conditions during transverse data taking:
Use laser #6, Vmod 1.2 Vpp, sweep/wait time = 6/30 or 8/30, (I did not see big difference between two sweep time settings); Use 1st order fit and calibration constant for "one laser and cooling jet on".
details on EPR data:
1st order fit.
down: freq. shift = -38.97 KHz +/- 1.25 KHz, P = -39.56%+/- 1.26%
up: freq. shift = 38.65 KHz +/- 1.23 KHz, P = 39.22%+/- 1.24%