• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name kramerk,lingyan

    Log entry time 05:19:34 on September 7,2001

    Entry number 73664

    keyword=Target - NMR/EPR - 45.4%/44.4%

    NMR and EPR performed with 3 lasers on, gradient coils at (2.0L, 5.0T) in pick-up coils position.

    NMR polarization from LabVIEW - 45.4%
    EPR polarization from Xiaochao's code (and new calib. const.):
    0th order fit:
    down: freq. shift = -45.19 KHz +/- 1.19 KHz, P = -45.06%+/- 1.18%
    up: freq. shift = 44.70 KHz +/- 1.31 KHz, P = 44.58%+/- 1.31%
    1st order fit:
    down: freq. shift = -44.89 KHz +/- 1.10 KHz, P = -44.76%+/- 1.10%
    up: freq. shift = 43.48 KHz +/- 1.23 KHz, P = 43.35%+/- 1.23%

    comments: EPR works very nice with 3 lasers !