Log entry time 01:59:52 on September10,2001
Entry number 74029
keyword=Target - NMR/EPR - 38.9%/38.3%
NMR and EPR performed at around 01:20. Both were performed within the pick-up coils at 0 deg. conf. with the gradient coils @ (2.0L,5.0T). NMR was performed with no lasers on. EPR was performed with 3 lasers on.
NMR polarization extracted from LabVIEW - 38.9%
EPR polarization extracted from Xiaochao's code:
1st order:
down: freq. shift = -38.51 KHz +/- 1.21 KHz, P = -38.40%+/- 1.20%
up: freq. shift = 38.22 KHz +/- 1.23 KHz, P = 38.11%+/- 1.22%
comments : good flip, lots of field drift . . .