    User name Aniol

    Log entry time 23:59:04 on November11,2001

    Entry number 77038

    keyword=shift summary

    1600-1630,
    Trouble with coda. Finally got it back to running.
    Q2 on Left spec tripped off. Reset.

    1630-2000
    data taking at kin h- at ~68 uA
    20:09 start Iscan, 5uA run 2064
    20:23 start 10uA run 2065
    20:40 start 20 uA run 2066
    20:49 start 30uA run 2067
    20:58 start 45 uA run 2068

    run #_I(uA)__Q(mC)___T1____T3____T5_____T5/T1___T1/Q___T3/Q___T5/Q
    2064__4.68_3.437____1.336e7_0.8669e7_13106_9.81e-4_3.89e6__2.52e6__3813
    2065_10.2__5.993____2.310___1.501____37378_1.62e-3_3.85e6__2.50e6__6237
    2066_19.2__7.479____2.850___1.865____76368_2.68e-3_3.81e6__2.49e6__10211
    2067_23.6*_7.833____2.958___1.947___120280_4.07e-3_3.78e6__2.49e6__15356
    2068_35**_10.53_____3.933___2.607___234307_5.96e-3_3.74e6__2.48e6__22251
    2063_67.6_46.78_____17.25___11.57__1549593_8.98e-3__3.69e6__2.47e6_33125
    *beam off, actual current ~29 uA
    **beam off, actual current ~44 uA
    A plot of T5/T1 vs I is almost a perfect straight line, as expected. By eye, from the plot, it looks like rate(TRUE)/rate(T1) ~ 1.e-4. Comparison of T5 rate and T1 and T3 rates at 68 uA is consistent with a resolving time of 55ns.

    21:19 start run 2070, trigger efficiency run, all ps = 1
    21:24, reset ps to proper running values, start run 2071
    23:28 rf trip, accel. off to do a QE scan
    23:40 beam back
    great beam delivery