Main INDEX
Monthly INDEX
PREV
NEXT
User name M Jones
Log entry time 05:55:55 on May30,2002
Entry number 4244
keyword=Modified current calibration constants
Modified current calibration constants. In the window for the BCM epics I modified
that current calibrations to 77.40 and 79.06 for the upstream and downstream to
reflect the current calibration runs of May 6th.
I also modified the values for the calibration of the V-to-F scalers to beam current in the filehalla_scaler_process.tcl so that the currents and charges in the end-of-run will reflect the new calibration.