Main INDEX
Monthly INDEX
PREV
NEXT
User name Hyde-Wright
Log entry time 16:50:06 on November 7,2002
Entry number 89809
Followups:
keyword=DAQ
Run 3061, 3062 kin Q3_i40
Data rate=320 Hz, but Dead Time calculation is 9-10%.
Earlier runs at this rate had only 1% deadtime.
T9 is a 30 Hz trigger, which does not have a scaler, but
DAQ assumes at 30 Hz rate in computation of Deadtime.
Run 3061 replay shows NO T9 events (contrast with e.g. 3051).
Probably T9 is accidentally disconnected.