Main INDEX
Monthly INDEX
PREV
NEXT
User name paschke
Log entry time 11:43:19 on January 7,2003
Entry number 90766
keyword=IA feedback system not working
I attempted to perform feedback. I had to fake the IA slope, which currently is too
small (80 ppm/volt) to zero our large asymmetry (-1500 ppm). I plugged in a slope
of 800 ppm/volt, just to test the system.
The system failed this test. A charge asymmetry was measured and sent to the
parity lock server, all three switches for this system were "enabled", but the
timestamp was never updated and the DAC setpoint never changed.
FIGURE 1