User name paschke/kaufman
Log entry time 19:06:52 on January 8,2003
Entry number 91054
This entry is a followup to: 91025
keyword=IA system re-calibration
Beam is back, so we are continuing the IA re-calibration procedure.
17:42 We have coordinated with Hall C, and are starting our baseline
Run IA S2 S1 Aq(bcm1) Aq(bcm2) Starttime
1121 5 0 0 526 545 17:42
1122 9 0 0 929 971 17:49
1123 9 -1000 0 1096 1138 18:07
1124 9 -2000 0 1032 1073 18:19
.... charge asymmetry RMS went up, for no good reason.
1125 9 -2000 0 1006 1050 18:27
... charge asymmetry RMS went back down..
1126 9 -3000 0 799 829 18:34
1127 9 -1500 0 1032 1073 18:43
1128 9 0 0 1077 1122 18:48
1129 5 0 0 535 554 18:51
1130 1 0 0 -30 -36
1131 5 0 0 511 532
Ok, so things aren't as easy as I thought. The effect rolls over, and we can't get
much more than 120 ppm/Volt. Also, there is a lot of backlash, so much that
the amount of backlash approximately equals the effective moment arm. Hmmm..
this will take some more thinking.
We are losing beam (target stuff) so we are packing it in now. I'll tell MCC to keep the instructions around, in case someone on owl gets ambitious. We can use our
new IA slope to try feedback, as long as the "natural" Aq isn't too large. I expect that with so little breathing room around the systematic offset, we will be likely to rail...