    Log entry time 13:49:20 on January14,2003

    Entry number 91813

    keyword=Compton

    The two BPM's in the compton chicane read by coda developed problems.
    We suspected the sample and hold modules so,
    Tom Powers sent Pete Francis for trouble shooting.

    Problem was traced to a bad V-to-F module in the compton DAQ and shorting of wires in a junction box. After straightening out wiring and shiching V-to-F,
    The BPM readouts work fine now.